基于噪声等效辐射亮度的红外高光谱遥测系统辐射定标方法

Radiometric Calibration Method for Infrared Hyperspectral Telemetry System Based on Noise Equivalent Spectral Radiance

  • 摘要: 红外辐射定标是建立成像光谱的数字量化值与目标光谱辐射亮度值二者之间关系的方法,对所测光谱进行定标是定量分析中的重要环节。在实际应用中,由于仪器本身的稳定性和目标辐射特性对仪器的影响,定标结果的好坏没有很好的评估,本文提出了基于噪声等效辐亮度(noise equivalent spectral radiance,NESR)来定量分析定标结果的准确性。采用两点线性辐射定标法对仪器进行标定,利用NESR信噪比计算方法得到仪器NESR的实测值,通过对比NESR标称参数,可以快速、定量化地判定辐射定标结果的准确性。利用自研的成像遥测系统采集到的数据进行验证,得出该方法可以快速、有效地评价辐射定标结果的准确性,从而提高标定后亮温光谱的质量,为后续红外光谱气体识别奠定基础。

     

    Abstract: Infrared radiometric calibration is a method for establishing a relationship between the digital quantization values of imaging spectra and the spectral radiance values of a target. Calibrating the measured spectra is crucial in quantitative analysis. In practical applications, because of the stability of the instrument and the effect of the target radiation characteristics on the instrument, the quality of calibration results cannot be evaluated effectively. This study proposes a method based on noise equivalent spectral radiance (NESR) to quantitatively analyze the accuracy of calibration results. A two-point linear radiometric calibration method is used to calibrate the instrument, and the measured NESR value is obtained using the NESR signal-to-noise ratio calculation method. By comparing the measured NESR with the nominal NESR parameter, the accuracy of the radiometric calibration results can be determined rapidly and quantitatively. Data obtained using a self-developed imaging telemetry system are used for verification, which demonstrates that the proposed method can rapidly and effectively evaluate the accuracy of radiometric calibration results, thereby improving the quality of brightness temperature spectra after calibration and establishing a foundation for subsequent infrared spectral gas identification.

     

/

返回文章
返回