施德恒, 王栋臣, 黄国庆, 刘玉芳. 一种实用化实时测温系统测温精度的进一步分析[J]. 红外技术, 2005, 27(6): 433-438. DOI: 10.3969/j.issn.1001-8891.2005.06.001
引用本文: 施德恒, 王栋臣, 黄国庆, 刘玉芳. 一种实用化实时测温系统测温精度的进一步分析[J]. 红外技术, 2005, 27(6): 433-438. DOI: 10.3969/j.issn.1001-8891.2005.06.001
SHI De-heng, WANG Dong-chen, HUANG Guo-qing, LIU Yu-fang. Further Analyses about Temperature Measurement Accuracy of a Practical Real-time Temperature Measurement System[J]. Infrared Technology , 2005, 27(6): 433-438. DOI: 10.3969/j.issn.1001-8891.2005.06.001
Citation: SHI De-heng, WANG Dong-chen, HUANG Guo-qing, LIU Yu-fang. Further Analyses about Temperature Measurement Accuracy of a Practical Real-time Temperature Measurement System[J]. Infrared Technology , 2005, 27(6): 433-438. DOI: 10.3969/j.issn.1001-8891.2005.06.001

一种实用化实时测温系统测温精度的进一步分析

Further Analyses about Temperature Measurement Accuracy of a Practical Real-time Temperature Measurement System

  • 摘要: 基于Kirchhoff定律,利用半导体激光器InGaAs/I及钽酸锂热释电探测器设计了一种实用化的实时测温系统.基于反射辐射对系统测温精度影响的分析,确定了水冷遮蔽板的尺寸(即H/R之值);通过对A/D转换器件的分辨率对系统测温精度的影响及对V(T)-T曲线温度灵敏度制约的研究,确定了系统应选用16位的A/D转换器件;依照光路中的外界干扰(选择性吸收气体、烟尘等微粒)对系统测温精度影响的讨论,提出了其相应的抑制措施;总结出了影响系统测温精度的5种因素,并对其中的2种进行了简要的分析.该系统的测温不确定度优于0.3%.

     

/

返回文章